
mA OUT, % OUT (Percent of 4 to 20 mA)
Choose this function to provide an output from 0.00 to 24.00 milliamps. The compliance voltage is a
nominal 24 VDC to provide the driving power to your milliamp receivers.
1) Disconnect one or both input wires from
the device to be calibrated.
2) Select “mA” or “% 4 to 20mA” with slide
switch
q.
3) Select “SOURCE” using slide switch w.
4) Connect the output leads of the Meriam
Model M334 to the inputs of the device
being calibrated, making sure to check
polarity. Red lead to the plus (+) input
and black lead to the minus (-) input.
The output is adjusted by turning knob
r
while the EZ-CHECK™ switch e is in the
“DIAL” position, or the current can be set
at the fixed points of 4.00mA (0.0%) or
20.00mA (100.0%) with switch
e.
READ mA, READ % (Percent of 4 to 20 mA)
Choose this function to measure from 0.00 to +52.00 milliamps or -25.0 to 300.0%.
1) Open the current loop at any conve-
nient point along the signal path.
2) Select “mA” or “% 4 to 20mA” with
slide switch
q.
3) Select “Read” using slide switch w.
4) Connect the red input lead (+) of the
Meriam Model M334 to the more
positive point of the break and the
black input.
Signals below 0 mA or open circuits are
indicated by 0.00 mA (-25.0%) on the
display. Signals above 52 mA are current
limited by protection circuitry.
Sourcing Milliamps
Reading Milliamp Outputs
Meriam Process Technologies
10920 Madison
Ave.
• Cleveland,
OH 44102 Tel:
216.281.1100 • F
ax:
216.281.0228 •
[email protected] • www.meriam.com
Comentarios a estos manuales